This
analyzer can be configured to measure at any wavelength within the ultraviolet
or visible portions of the electromagnetic energyspectrum. Since numerous gases
and liquids have a useful absorbance peak in the ultraviolet region, a wide
variety of applications are possible. The analyzers optical path (figure
1) consists of a UV source, sample cell, and detector. A filter wheel located
in front of the detector continuously rotates a reference and measuring filter
in and out of the optical path. The material of interest continuously flows
through the sample cell where the optical energy is absorbed at the various
wavelengths in accordance with its composition. The analyzer utilizes the
reference wavelength to compensate for extraneous phenomena (turbidity, dirt on
optics, source drift, etc.) thereby producing a high degree of stability over
long periods.